Evaluating and benchmarking the EEG signal quality of high-density, dry MXene-based electrode arrays against gelled Ag/AgCl electrodes.
Erickson B
,
Rich R
,
Shankar S
,
Kim B
,
Driscoll N
,
Mentzelopoulos G
,
Fernandez-Nuñez G
,
Vitale F
,
Medaglia JD
.
???displayArticle.abstract???
Objective.To evaluate the signal quality of dry MXene-based electrode arrays (also termed 'MXtrodes') for electroencephalographic (EEG) recordings where gelled Ag/AgCl electrodes are a standard.Approach.We placed 4 × 4 MXtrode arrays and gelled Ag/AgCl electrodes on different scalp locations. The scalp was cleaned with alcohol and rewetted with saline before application. We recorded from both electrode types simultaneously while participants performed a vigilance task.Main results.The root mean squared amplitude of MXtrodes was slightly higher than that of Ag/AgCl electrodes (.24-1.94 uV). Most MXtrode pairs had slightly lower broadband spectral coherence (.05 to .1 dB) and Delta- and Theta-band timeseries correlation (.05 to .1 units) compared to the Ag/AgCl pair (p< .001). However, the magnitude of correlation and coherence was high across both electrode types. Beta-band timeseries correlation and spectral coherence were higher between neighboring MXtrodes in the array (.81 to .84 units) than between any other pair (.70 to .75 units). This result suggests the close spacing of the nearest MXtrodes (3 mm) more densely sampled high spatial-frequency topographies. Event-related potentials were more similar between MXtrodes (ρ⩾ .95) than equally spaced Ag/AgCl electrodes (ρ⩽ .77,p< .001). Dry MXtrode impedance (x̄= 5.15 KΩ cm2) was higher and more variable than gelled Ag/AgCl electrodes (x̄= 1.21 KΩ cm2,p< .001). EEG was also recorded on the scalp across diverse hair types.Significance.Dry MXene-based electrodes record EEG at a quality comparable to conventional gelled Ag/AgCl while requiring minimal scalp preparation and no gel. MXtrodes can record independent signals at a spatial density four times higher than conventional electrodes, including through hair, thus opening novel opportunities for research and clinical applications that could benefit from dry and higher-density configurations.
???displayArticle.pubmedLink???
38081060
???displayArticle.link???J Neural Eng ???displayArticle.grants???[+]
Alba,
Novel hydrogel-based preparation-free EEG electrode.
2010,
Pubmed
Alschuler,
Identifying electrode bridging from electrical distance distributions: a survey of publicly-available EEG data using a new method.
2014,
Pubmed
Bazanova,
Interpreting EEG alpha activity.
2014,
Pubmed
Burgess,
How reproducible is the topographical distribution of EEG amplitude?
1997,
Pubmed
Chu,
High density EEG-what do we have to lose?
2015,
Pubmed
Delorme,
EEGLAB: an open source toolbox for analysis of single-trial EEG dynamics including independent component analysis.
2004,
Pubmed
Driscoll,
MXene-infused bioelectronic interfaces for multiscale electrophysiology and stimulation.
2021,
Pubmed
Erickson,
Evaluating and benchmarking the EEG signal quality of high-density, dry MXene-based electrode arrays against gelled Ag/AgCl electrodes.
2024,
Pubmed
,
Echinobase
Fan,
Testing the efficiency and independence of attentional networks.
2002,
Pubmed
Fiedler,
Contact Pressure and Flexibility of Multipin Dry EEG Electrodes.
2018,
Pubmed
Gevins,
Spatiotemporal dynamics of component processes in human working memory.
1993,
Pubmed
Guger,
Comparison of dry and gel based electrodes for p300 brain-computer interfaces.
2012,
Pubmed
He,
Scale-free brain activity: past, present, and future.
2014,
Pubmed
Hinrichs,
Comparison between a wireless dry electrode EEG system with a conventional wired wet electrode EEG system for clinical applications.
2020,
Pubmed
Hsieh,
Design of hydrogel-based wearable EEG electrodes for medical applications.
2022,
Pubmed
Iivanainen,
Spatial sampling of MEG and EEG based on generalized spatial-frequency analysis and optimal design.
2021,
Pubmed
Kam,
Default network and frontoparietal control network theta connectivity supports internal attention.
2019,
Pubmed
Kanske,
Attentional orienting towards emotion: P2 and N400 ERP effects.
2011,
Pubmed
Kappenman,
The effects of electrode impedance on data quality and statistical significance in ERP recordings.
2010,
Pubmed
Ko,
Flexible graphene/GO electrode for gel-free EEG.
2021,
Pubmed
Krachunov,
3D Printed Dry EEG Electrodes.
2016,
Pubmed
Kuhnke,
High Frequency Oscillations in the Ripple Band (80-250 Hz) in Scalp EEG: Higher Density of Electrodes Allows for Better Localization of the Seizure Onset Zone.
2018,
Pubmed
Kübler,
The user-centered design as novel perspective for evaluating the usability of BCI-controlled applications.
2014,
Pubmed
Lachaux,
Measuring phase synchrony in brain signals.
1999,
Pubmed
Lantz,
Epileptic source localization with high density EEG: how many electrodes are needed?
2003,
Pubmed
Li,
Review of semi-dry electrodes for EEG recording.
2020,
Pubmed
Li,
Polyvinyl alcohol/polyacrylamide double-network hydrogel-based semi-dry electrodes for robust electroencephalography recording at hairy scalp for noninvasive brain-computer interfaces.
2023,
Pubmed
Li,
Towards real-life EEG applications: novel superporous hydrogel-based semi-dry EEG electrodes enabling automatically 'charge-discharge' electrolyte.
2021,
Pubmed
Lin,
Novel dry polymer foam electrodes for long-term EEG measurement.
2011,
Pubmed
Lopez-Calderon,
ERPLAB: an open-source toolbox for the analysis of event-related potentials.
2014,
Pubmed
Lopez-Gordo,
Dry EEG electrodes.
2014,
Pubmed
Mentzelopoulos,
Alerting attention is sufficient to induce a phase-dependent behavior that can be predicted by frontal EEG.
2023,
Pubmed
,
Echinobase
Michel,
EEG Source Imaging: A Practical Review of the Analysis Steps.
2019,
Pubmed
Michel,
Towards the utilization of EEG as a brain imaging tool.
2012,
Pubmed
Murphy,
A gel-free Ti3C2Tx-based electrode array for high-density, high-resolution surface electromyography.
2020,
Pubmed
Murphy,
Time Evolution of the Skin-Electrode Interface Impedance under Different Skin Treatments.
2021,
Pubmed
Muthukumaraswamy,
High-frequency brain activity and muscle artifacts in MEG/EEG: a review and recommendations.
2013,
Pubmed
Nemtsas,
Source localization of ictal epileptic activity based on high-density scalp EEG data.
2017,
Pubmed
Odabaee,
Spatial patterning of the neonatal EEG suggests a need for a high number of electrodes.
2013,
Pubmed
Oliveira,
Proposing Metrics for Benchmarking Novel EEG Technologies Towards Real-World Measurements.
2016,
Pubmed
Oostenveld,
The five percent electrode system for high-resolution EEG and ERP measurements.
2001,
Pubmed
Peirce,
PsychoPy2: Experiments in behavior made easy.
2019,
Pubmed
Petrov,
Ultra-dense EEG sampling results in two-fold increase of functional brain information.
2014,
Pubmed
Pourahmad,
Evaluation of a Low-cost and Low-noise Active Dry Electrode for Long-term Biopotential Recording.
2016,
Pubmed
Puce,
A Review of Issues Related to Data Acquisition and Analysis in EEG/MEG Studies.
2017,
Pubmed
Radüntz,
Signal Quality Evaluation of Emerging EEG Devices.
2018,
Pubmed
Robinson,
Very high density EEG elucidates spatiotemporal aspects of early visual processing.
2017,
Pubmed
Rusiniak,
The relationship between alpha burst activity and the default mode network.
2018,
Pubmed
Seeber,
Subcortical electrophysiological activity is detectable with high-density EEG source imaging.
2019,
Pubmed
Smith,
Assessing and conceptualizing frontal EEG asymmetry: An updated primer on recording, processing, analyzing, and interpreting frontal alpha asymmetry.
2017,
Pubmed
Widmann,
Digital filter design for electrophysiological data--a practical approach.
2015,
Pubmed
Zelmann,
Scalp EEG is not a blur: it can see high frequency oscillations although their generators are small.
2014,
Pubmed
Zrenner,
The shaky ground truth of real-time phase estimation.
2020,
Pubmed